News Categories
Sorry, you do not have permission to carry out this action.
Latest Updates
Technical Support Options
Posted by ADMIN - Vale on 19 July 2023 14:03

Prior to contacting Channel Assist, please search these self-serve options for a faster answer to your question.

Field Portal

 (Agilent Internals, ASP and IDOs)

  • Manuals
  • Part Numbers
  • Service Notes
  • SPIFM Documents
  • Troubleshooting Tools

IKB Agilent Internal Knowledge Base

 (Agilent Internals and ASP)

Troubleshooting information, tips & techniques, created by Agilent engineers based on real issues, reviewed by experts from Channel Assist (GSCA), Call Centers (GSCC) and Solution


Read more

For Agilent Employees, Authorised Partners and Distributors Only
Posted by Patrick Barrett on 28 October 2015 17:39

Instructions on how to obtain access to the support centre

Access to our support is by user name and password. the user name will always be your email account.

We accept email accounts from the following:

Agilent employees Your @agilent.com email account
International distributors An email account from your company domain
Authorised Support Partners The email account registered with your Agilent manager

We will also ask that you identify your Agilent appointed manager in case we


Read more


Agilent Technologies - Global Service Channel Assist

 News Categories
knowledgebase
Hi all,

There is a service note about MCGV failing the gradient test G711X/G5654 Pumps-008A.

We know that some MCGV below serial number 1200000 can fail the gradient test, but it doesn't mean that automatically the MCGV must be replaced with a new one with SN above 1200000. Please run the qualification with the old MCGV and follow the procedure in the attachment of the service note. If after 2 or 3 attempts the old MCGV is not passing the gradient test, it can be replaced with a new one.

All the logistic centers only store the new type of MCGV with SN 12xxxxx, so if you order a new MCGV, you should only receive the latest version.
Many customer's have the need to collect data directly to a network drive. Collecting data to a network drive is usually not a good idea if any issues arise with the
network hardware or software the raw data can be lost. The preferred way is to complete this task is to collect the data on the local file system and then copy to the network location after the run completes.

To setup post run file copy you can use a macro called "nas_copy.mac" to copy the file from the local drive to network driver AFTER the data is collected.

"nas_copy.mac" is a CONTRIBUTED UTILITY to be used for the purpose
of exporting data files (and associated directory structure) acquired on
Agilent MassHunter GCMS Acquisition platforms to networked or secondary drives.

The remote file system MUST have directories created for the file transfer before
the configuration process will allow destination selection. The directories MUST
be named with a .NET extent (i.e. msd1.net, gcfid3.net, etc.). Once the destination
!directories have been created, use the ConfigCopy macro to select them for use
with the instrument. From the command line in MH GCMS Acquisition, type the following:

macro "NAS_Copy.mac" <enter>
ConfigCopy <enter>

Follow the prompts on the screen to select the destination for data files. This
will allow multiple instruments to share the same drive and store the data in
individual directories. The configuration information is stored in the file
nas_copy.ini, located in the instrument folder (i.e. masshunter\gcms\x where x=instrument number).
This allows multiple instruments on the same PC to send data to unique locations.

To automate the file transfer, edit your MH GCMS Acquisition Method and add the following
command in the Method Information Post-Run Macro/Command field:

NAS_Copy
Oct 2017: Ion Source Temperature in NCI mode for SQ, QQQ and QTOF

Issue: Turning on the filament will affect the Ion Source temperature.

If acquiring with the default IS temperature of 150°C in NCI mode by the end of the run the IS will be increased.
This might be up to 20°C higher so for the next run there will be a lay waiting to get the IS back to 150°C again.

Workaround, run the IS at 160°C and the Quad at 106°C instead.

Below are the test results running at different temperatures:


______________________________________________________________

GPD Support Channel Assist
Dec. 2017              Part number of sapphire piston for 1290 pumps

We had several cases reported shorten life time issue of 1290 pump heads while MeOH (Methanol) with acid (regardless which type of acid) in use. Due to more often customer use ACN (Acetonitrile) as organic solvent in reverse phase analysis and pure MeOH doesn't show any shorten life time issue, so root cause remaining unknown for us.

The symptom is ZrO2 piston showing scratch in short period of time by using MeOH with acid as mobile phase. Same application using in conventional LC which equipped with sapphire piston has no such issue.

Anyway, we do have sapphire plunger for 1290 pump, part number is 5067-6174 1ea.

This P/N could be use in 1290 Classic and EM (Easy Maintenance) pump head but not LL (Long Life) pump head.

FYI.
Jan 2018       

FSE reporting that new G7102A ELSD is use LSU as response unit which is different with old G426X A/B are use mV as response unit.

Here is the reply from SU.

There's only G7102A HDR (High Dynamic Range) use LSU but not for G426X A/B. The G426X A/B will always have the mV and the G7102As will always have the LSU. Both units are not convertible.

Up to a value of ~ 300 LSU both the G7102 and G426X A/B are the same …. Basically 1mV = 1 LSU up to 300 ….

What’s the orderable part number for the 1260 G1312B SSV?

The update kit G1381-6000 seems to be obsoleted.


PSE Tobias informed that

The pure SSV has the part number G1312-60068 for G1312B and 5067-5895 for G7112B.




Problem:

The methane flow adjument fails and autotune can not be started. 

Cause:
Before the CI autotune is performed, the system will optimize the flow of methane into the source to adjust the pressure to the optimum level.

The system is looking for a target ration of m/z 28 and m/z 27.

Mass 27 = C2H3+
Mass 28 = C2H4+

(See below for complete list of reactions that occur with methane in the source).

The ratio of the intensity of Mass 28 / Mass 27 is used to adjust the correct pressure in the trap. Default target values are 2.2 - 2.3 



Solution:

If the system can not reach a target of 2.2-2.3 more then likely there is a leak somewhere. 

 The usual culprit is the seal between the transfer line and the ion source.  Below is more detailed list of possible problems.

  1. Seal between the transfer line and the ion source.  make sure the Tip seal and spring are correctly installed.
  2. Corrupt tune file - reload a default pcich4.u tune from the mssetup directory.  Go to manual tune and manually change all quad parameters you have a working EI tune for.
  3. No reagent gas flow:  check the monitor
  4. Wrong reagent gas
  5. Wrong ION Source or mis-assembled ion source.
  6. Sideboard connectors, not connected.
  7. Go back to EI an make sure EI tune acceptable 
  8. Make sure EI air/water values are OK 
More information on reactions in source for methane CI reagent.


https://www.agilent.com/cs/library/primers/public/GC_handbuch_E.pd
 

24-Aug-2018

Waters special detector (RID, ELSD & FLD) IPBs are available on support portal

http://intranet.chem.agilent.com/sites/partners/support/Product%20Pages/Special%20Detectors.aspx


Sep 2018
Introduction:
In order to downgrade a 5977 from HES to Extractor ion source it is basically needed to replace the Source, the Source Radiator and the Ceramic Board.

Procedure:
The needed parts are:

G7000-60827 Wire, Extractor Lens
G7000-67625 Source PCA/Cable Assy GC/MS QQQ Ext
G7002-67037 Source Radiator Assy - Packaged
G3870-67720  Extraction Source Assy,  (with short entrance lens)

The extractor ion source has the short or standard entrance lens installed.
So do not mix with the Estractor Source for a 7000 unit, it would be the 
G7003-67720 with long entrance lens.
PN for the LVPS for the 2475 is 700001641 and is the same for both the platinum and blue versions of the detector.
There are three different PN for the QSM degasser chambers and they are as follows:

MV-W-700005172: Single degasser chamber (standard) -- THIS HAS BEEN OBSOLETED AND REPLACED WITH MV-W-700011667
MV-W-700002602: complete degasser assembly
MV-W-700011667: stented degasser chamber used for GPC/normal phase solvents; this is also the degasser chamber used on the QSM PLUS

cooling is not available with G1364E collector.

15711720814c8fab892119479e667063f559cb6d45ffaeb0c7.png

4 May 2018
Q: How to turn off the UV lamp after run in OpenLab CDS 2.X?
A:
  1. Close acquisition client if the acquisition client is working.
  2. Close connection
  3. On control panel, click Administration > instruments options
  4. Select turn off detector > enter 2 minutes,
  5. Click save settings
  6. Launch the instrument (open the acquisition client).
  7. Run sequence or single sample analysis
  8. Wait for the run queue is idle (no pending items in the run queue)
  9. And then wait for 2 minutes (set in step 4)
  10. The lamp should turn off (will delay about 0.1 minute)

 

29-May-2018

Q: Abnormal DC voltage from power supply of G1315C caused communication failure with PC. 
A:  Recently, we got one case, saying that G1315C can not talk to Chemstation or Labadvisor. The communication can not be established even by Ping 192.168.254.11. Normally, FSE would think it is the problem of the detector mainboard. However, in this case, the communication still did not succeed after replacing with a new mainboard. Later, the DC voltage supplied by module's power supply were checked. Among various DC values provided by power supply, the 24VDC is suspicious, the real measured one is merely 21.3V. Then, we changed a new power supply and brought the 24V to 23.6V. As a result, the communication was restored. 
      So, when coming with communication problem, we should not only doubt on mainboard, but also power supply. Below is the correct reference voltage values for G1315C

RF Power supply troubleshooting:

Test RF Power supply to 150V from diagnostics (remember first click gives 50V and then another click gives 150V).
If NO voltage is observed then isolate the RF power supply and SSRF following next steps:
1.Ensure RF power supply is off
2.Disconnect RF power cable from Filter board. (2 pin Red/Grey wires connector, lower connector on the RF filter board on the rear, this is the RFPS feed up to the SSRF)
3.Ensure interlocks are satisfied (RF door closed)
4.Set 50V and 150V RF power supply (click twice to get to 150V) on Diagnostics > Instrument > Plasma section
a. Can you hear the relay click in the MPM?
b. can you see the RFPS power on and the fans start to rotate?
If this happens look at the voltage on the diagnostics, it should be approx. 50v on the left side and ( with the cable disconnected) approx. 0v on the right side. 
If not voltages reported then RFPS or MPM are faulty, or there is an interlock problem, or connection for the RFPS is faulty.
Recommended parts:
Assy RF Power Supply G8010-87301
RF power supply connector cable assy G8010-60801

SSRF Troubleshooting:

If RF power supply voltages are ok then this indicates that the RF power supply is able to generate the voltage but is being loaded by something in the RF Box, in this case:
1. Switch off the RFPS, refit the RF power cable to the Filter board.
2. Remove SSRF cover and disconnect electrical cable, close door and enable 150V.  if Reported and Measured voltages are OK then the RF power supply is loading the and SSRF is faulty.
3. If measured voltage is not OK with the SSRF disconnected then Filter board or SSRF control board are faulty.
Recommended parts:
Assy PCA ICP SSRF Filters G8010-65207
Assy PCA ICP SSRF Control G8010-65205
SSRF Assy G8010-60610

Oct 23rd
Q: How to check which fan is in error state for LC 1120/1220?
A: For “Fan errors” in 1220/1120, it is always confusing. When a sub-module got problems in the system, you will receive “fan errors” for all sub-modules, even for example the oven does not have a fan. But there are some clues which is indicated by error code. In you case, if you saw error code “68, 1”, then it refers to the fan in VWD module. If you saw “68.0”, then the bad guy should be the fan in pump module.
Hi all,

GPC SEC software is compatible with windows 10 64 bit and Windows 8.1 (32 & 64 bit) , but the PC must have the latest windows updates installed and the Agilent GPC/SEC A.02.01 SR3 Instrument Driver version. The drivers are available in the support portal in the link below:

http://intranet.chem.agilent.com/sites/partners/support/Product%20Pages/GPC-SEC%20software.aspx

27th Jan 2018
Q: What is the difference between power switch board 5067-6140 and 5067-5712?
A: 5067-6140 is the power switch board with coating, this is especially used for G7129X and G7167X with cooler. If cooler is used, there might be a risk of condensation water flowing to the location where the power switch situated. The coating of the board prevents short circuit of the board or other electronic components.
     5067-5712 is still used for other infinity II modules like pump and detector.

Migrate to the Internal knowledge Base https://community.agilent.com/docs/DOC-10368 

Hello,

There are currently 3 options available for the customers. Below there is a clear comparison of the features for each platform. As you can see the OpenLab CDS is the most complete option and the only limitation is that it doesn´t support multi detection. Viscometer and Light Scatering=LS (please do not confuse it with Evaporative Light Scatering or ELSD) are not supported in OpenLab CDS.

Hi Annelies:
     This is a known issue. G1364-84545 is the new type tray only used for 1260 Inifinity II FRC. G1364-84525 is the old tray, which can be used for old and new FRC. 
     The new tray has different RFID coding, hence it is not supported by old 1100/1200/1260 FRC. Below is the table listing supported trays for different FRC



Best regards,



Steve Zhang

Global Service Channel Assist

----------------------------------------------------------------------------------------------------

Yes, 7100 lift tower(G7100-67009) is available for 1600.

NO NEED TO MIGRATE

After dicussions with product support this set up is untested and therefore unsupported.

Although the lift tower will physically fit problems in future may occur.
besides, the G1600A is obsolete and out of support.

From LA (FSE license) you can use the command line for the IP from Instrument Control.
Here you enter SER "DE12345678", click on SEND and check for (green) confirmation on the top.
Disconnect and do a physical reboot, so disconnect and reconnect the IP.
Now the new serial number should be shown in ‘detail screen.

 

No, G7112B could not execute the Solvent Compressibility Calibration.

This is the comment from the prodcut manager Jens, Hühner.  

My name is Jens and I am the backup product manager for the Lab Advisor. I checked your issue with the Solvent Compressibility Calibration.
This  calibration was only implemented in the firmware for the G1312B Module (the old 1260 Infinity I Bin Pump).
Therefore, the calibration is also only available in Lab Advisor for this specific module.
I think it is a mistake that the calibration, as a feature, is included in the manual for the new 1260 Infinity II Bin Pump (G7112B).
The Solvent Compressibility Calibration is definitely not implemented in the firmware for the 1260 Infinity II Bin Pump (G7112B)
and therefore also not available in Lab Advisor. Unfortunately this is no bug in the Lab Advisor.

Part number is 5065-4493 OQ/PV capillary for prep LC system, SS

24th Dec 2017
Provide a part list for G7112B
Hello Thomas,

PC recommendations is
16GB for Windows 10 Pro 64-bit.

Please check the attached document to get more.
I also attached Windows firewall configuration.
Agilent doesn't recommend to use antivirus software because they don't install any antivirus softeware at software test.
But I know some cusgtomers who use antivirus softare with MH software.



with best regards,



Masataka Adachi

______________________________________________________________

GPD Channel Assist
Hello,

P/N is G7003-60800 ( LED Cable ).
Note, the connection is Mainboard 2 Status. Not Mainboard 1.

with best regards,



Masataka Adachi

______________________________________________________________

GPD Channel Assist
Agilent does not offer a re-certification service for the Starna kits or other UV/Vis OQ standards and filter. These kits are managed locally.

The options are to either send the kits directly to the manufacturer for re certification or replace the kit with a new one. 

Please consult with your district manager or Agilent liaison to determine  the local procedure.
Kit Water Tubing PN G8010-67096

This Kit of internal water tubing only include the four tubings (Red, Green, Yellow and Polyurethane).

If ferrule and brass inserts are also required then following parts must be ordered separately:

1610119800 Ferrule -6mm OD Tube- Swagelok (Qty = 1)
1510231600 INSERT-SWAGELOK FOR 6MM OD X 4MM ID TUBE (Qty = 1)
Hello Kicki,

I have an update code for QTOF CI mode.
Please try to this code.


Here is an update PY code for the QTOF, the attached file gets copied to the C:\GCMS\msexe folder.



As a reminder, the issue is that with anything except a brand-new filament the emission is getting set to low when it is adjusted after the gain step in autotune.  The starting point for the emission adjustment is 2uA.  With a filament that has been used, even for just a few tunes or method runs, ions aren’t seen until the current reaches around 6uA. 



The was the code that I was looking at and I have seen better PCI and NCI tunes, if you can give it a try please.




with best regards,



Masataka Adachi

______________________________________________________________

GPD Channel Assist
Hello Jamie,

I received the feedback from Japanese GC Triple enginer.
He found a mistake in cable connection at smartcard in GC Triple Ject Clean unit.
Please check out the cable connection at rear.
And check out this servicve note , too.
http://intranet.chem.agilent.com/sites/partners/support/Service%20Notes/Service%20Note%20Viewer/MSD/GCMS7010/SN_GCMS7010S_012.htm


with best regards,



Masataka Adachi

______________________________________________________________

GPD Channel Assist
Hello Krzysztof,

What acquire mode does the customer ? Scan mode ? SIM mode ? SIM/Scan mode ?
If the customer take SIM/Scan mode, the loaded data file looks zig zag such like your attached.
The customer need to separate the data to SIM and Scan.
Go to "Extract Chromatograms...."
Select "TIC" in Type, select MS in MS level.
Then select "Scan" in Scans.
Click OK button.
TIC data will be extracted.

Select "TIC" in Type, select MS in MS level.
Then select "SIM" in Scans.
Click OK button.
SIM data will be extracted.






with best regards,



Masataka Adachi

______________________________________________________________

GPD Channel Assist
Jan 2018 MassHunter and position of Multi Postion Valve - MPV

To control the MPV from MassHunter acquisition you need to use a command to set the position.

Like: gcdirectget (“gc.mpv.5”,1)  - this will send the MPV to position #5.

Attached are some examples for the different valve positions.
Simply add them as a preprun macro.

Hope this helps.

Very best regards,
Max


______________________________________________________________

GPD Support Channel Assist

Please make a note that 3rd party support and info mailboxes no longer available and have moved to a support portal. Please submit your inquiries via the following portal:

 

https://www.agilent-labinformatics.com

 

If you do not have an account with this portal, you will need to perform a one-time registration by clicking on the Register button and answering a few questions. Upon the completion of the registration, a verification email will be sent to you and you will need to verify your email by clicking on the link in it. Once your email is verified, it will take up to one business day to activate your account and you will be able to submit your inquiries (tickets).

In case you are already registered with the portal, please submit your inquiry by pressing Submit Ticket button and selecting Third Party Software department
Hello Victor,

Please refer to the attached document.

with best regards,



Masataka Adachi

______________________________________________________________

GPD Channel Assist
from the ChemStation command line. 

print SendModule$(LAFC1,"tray:alig 0,0")

Hello Andreas,

Is it an Agilent Bundle PC ?

Before run sequence, check mem_workset. 
If mem_workset is greater than 450Mbytes, It will throw up an alert box, suggest that
the Acquisition be restarted.

To disable memory test, set the following entry into [PCS] section in C:\GCMS\Msexe.
 [PCS]
 NOMEMSTEST=1

with best regards,



Masataka Adachi

______________________________________________________________

GPD Channel Assist

13th Apr 2014
Q: Will the electromagnetic sending by phone transmitter station?
A: The baseline of an LC detector can be affected by mobile phone radiation. There are electromagnetic field immunity tests carried out on our modules and they´re compulsory, at least for Europe. Those tests are done with a field intensity of 3V/m resp. 1V/m.

 IEC/EN60004-3 Radiated Filed Immunity: 80MHz – 2.0GHz: 3V/m; 2GHz – 2.7GHz: 1V/m

03 May 2018
Q: What is the part number for the plug channel(air diffuser) on the top of air channel adapter for G1329X used with G1330B?
A:
    The part number is G1367-47200, it is in the accessory kit of G1330B.

29-May-2018
   Q:  What is the part number of the front kit for G7129X?
   A:  The door of G7129A can only be ordered as a pair, not a single one. The part number for door kit is G7129-68702
   
27-June-2018
Q: What is the capillary part numbers of injection valve-Metering for various autosamplers?
A:  G1329B                                     G1313-87301
     G1367E                                      G4226-60301
     G4226A                                      G4226-60301

     G7129A                                       5500-1228
     G7129B                                       5500-1228
     G7167A(strandard/multiwash)    5500-1159
     G7167B(strandard/multiwash)    5500-1159


July 3rd
Q: what is the part number for rotor seal of G1968F active splitter?
A:  Here the P/N for the G1968F active splitter and the G7170B MS Flow Modular which are identical.

0101-1393                          – Replacement Stator

1535-5215                          – RheBuild kit (3x stator and rotor seals)

22nd-Aug-2018
Q: What should we do after upgrading a G7110B without degasser with degasser G4280-60070?
Answer:

After installation,  the degasser option needs to be turned on by command: OPT:DGAS 1 in labadvisor

If the degasser needs to be removed, input command: OPT:DGAS:RESET

8th Sep 2018
Q: What is working principle of G7111X integrated degasser?
A: 

Two types of error form integrated degasser about reaching and maintaining the vacuum can be generated.

Error ID 8053 vacuum limit not reached;

Error ID 8014 vacuum cannot be maintained;

8053 is generated after pump start-up when the vacuum does not reach to the target default value 100mBar (+/-20mBar) within 8 min as a default timeout.

8014 is generated during continuous operation if the vacuum drops higher than 80mbar than target value – 100mbar as default.

The error 8053 can be caused by excessive solvent vapor concentration in vacuum chamber after long period of pump idle condition.

Evacuation mode initiated through the LabAdvisor can quickly help to resolve the situation in this case.

8th Sep 2018
Q: What is the part number of needle seat adapter for G4226, G1367 ALS?
A:
8th Sep 2018
Q: How to trouble shoot the extensive baseline noise of UV detector?
A: See attached document
8th Sep 2018
Q: How to handle the TFA application in LC?
A: See attached application document.
8th Sep 2018
Q: How to trouble shoot the ghost peak caused by gradient elution?
A: See attached document
10 Oct 2018
Q: What is the O-ring part number used for ELSD Nebulizer mount assembly?
A:  4---PL0890-0540
     5---PL0890-0545
Oct 2018
Introduction:
There have been some cases where new 7820A GC Logic Boards were shipped with a "blank" IP Address (i.e. 0.0.0.0), thus making not possible to connect to the GC and/or use the Remote Control Panel and set a valid IP Address.

Procedure:
Follow these steps to reset the IP Address to the default factory value:

1) Turn off the system;
2) Press and hold the [Prep Run] button on the GC front panel;
3) While holding the [Prep Run] button pressed, turn on the system, and wait for the initialization cycle to complete (until the "Agilent Technologies" logo appears on the display);
4) Release the [Prep Run] button;

The IP Address should have been now reset to default value. If needed, it is now possible to use the Remote Control Panel to connect to the system and change the IP Address to a different one.

#7820 #default #IP Address
Hi All,
I have attached a document "Ion Gauge Basic Principles of Operation", I have used information from the manufactures and ICP-MS service manual.

I hope this helps.

Cheers,
Glenn.

There is an issue when loading previous Verification and Diagnostics (V&D), Wavelength Accuracy reports.

B.05.02 behaviour: only the most recent results can be loaded.
Why?
UV-Vis Chemstation automatically saves a .VD and .VSD file for each result. The .VD file does not have a unique name, so each time a new V&D result is collected, the previous .VD file is overwritten. Since UV-Vis Chemstation can only load .VD files, only the most recent report can be opened.

B.05.04 behaviour: previous V&D results can be loaded, but the loaded report may display incorrect information.
Why?
UV-Vis Chemstation automatically saves a .VD and .VSD file for each result. The .VSD file does not have a unique name, so each time a new V&D result is collected, the previous .VSD file is overwritten. The .VSD file contains spectra, so the following error will appear:

UV-Vis Chemstation will display a report, but the report will contain incorrect or missing information. For example:

  • Spectra will not be displayed
  • Wavelength Reproducibility will not be displayed
  • Wavelength Accuracy table headers display "Found WL" instead of "Average WL"
  • Deviation values are inverted. For example, 0.10 will become -0.10.
  • Wavelength Accuracy table results are from a single collect instead of an average of ten collects.

Note: "Previous V&D" result refers to a result that is loaded after a the collection of a different result.

Workaround: After collecting Verification and Diagnostics results, save and/or print the reports.

Hi,

When you have a peak like below during the wavelength verification is worth repeating the test a couple of times, because with flushing is very likely the peak shape will improve. See attachment for your reference
NOV 2018

How to change valve type on G1316A?

Once field/customer decide upgrade valve to different type, due to valve itself has no RFID tag as 1290 TCC (G1316C) or MCT (G7116X). FSE has to manually change valve type after hardware installation.

The most easy way is using Lab Advisor FSE mode.

You could enter following command via command line under instrument control.

vt 1          represent 2 position 6 port valve
vt 2          represent 2 position 10 port valve

We are always encourage FSE turn TCC power off and on after received RA 0000. It helps on valve type stay on main board. 

If someone need to enter command via Chemstation, please refer to G1316-016 service note for the detail.
Hello Justin,

USP 1058 outlines general guidelines for AIQ (analytical instrument qualification) and the qualification life cycle. Details of instrument operating parameters and qualification requirements can be found in the USP general chapters for specific instrument types.The customer should have a qualification plan for the instrument that describes the intended use of the equipment and the tests required to ensure that it is operating properly.

USP 857 is a general chapter that contains specific guidelines for UV/Vis spectroscopy. Agilent's standard OQ protocols for UV/Vis meet the minimum requirements of USP 857 for most customers. Depending on the intended use of the instrument, the optional tests can be added to the EQP to create a custom EQP that meets the customer's unique needs. 

The attachment ('Using ACE to meet USP General Chapter 857') explains how the protocols in ACE can be used to meet USP 857 compliance. This document, provided by Marketing and R&D, can be found on Support Portal. Feel free to share this explanation document with the customer. Before the qualification is scheduled, the customer must approve the EQP review document and/or request a custom EQP.

Training materials (see the General Compliance and UV/Vis compliance with ACE CBT courses) contain topics that explain regulatory standards, the EQP review process and Agilent's UV/Vis compliance protocols. Certification for these courses must be updated annually. The Compliance Product Support team encourages the use of existing resources, but we are here to help if you still have questions or concerns.

Please review the explanation document and EQP with your customer. If the customer still has questions or concerns about the EQP, let us know!




Best Regards,



Colette Gay

SSD Product Support Engineer
28 Jan 2019
Q: What is the function of white ceramic plate on top of internal tray?
A: The ceramic plate on the top of internal tray is ONLY used for the G1364D Micro Fraction Collector to check the alignment of the capillary tip during initialization. Since we are using the internal tray for all G1364X Fraction Collectors this part is in, even only used by the Micro FC.
28 Jan 2019
Q: Why B pump head is responsible to build pressure when doing A channel leak rate test for G1312B?
      If we do the test for A channel, we were asked whether damper was included or not

If we choose no damper, then we found the pressure will be built up by solvent delivered with Channel B, instead of channel A

Once the target pressure reached, the channel A started to deliver and measure the leak rate.

If we choose include damper, then we found the pressure will be built up by solvent delivered with Channel A.

Once the target pressure reached, the channel A continued to deliver and  measure the leak rate.

Why the situation is that for “no damper”? Any benefits?

A: The reason for this is right positioning of the pistons before the test. In case of “without damper” the system is less “elastic” and it is more difficult to position the pistons in tested channel properly right after building up the pressure if it would be used for it.

March 13rd 2019
Q: FSE onsite got below error for a sample thermostat in G7129A intermittently. The thermostat can work and reaches at the set temperature point, but it will show below error after a certain period of time.

Module:             G7129A:DEAEQ02518 FW Rev: D.07.26 [0001]

Maximum Entries:    15

Time of last Entry: 2019-01-31 12:15:14

===============================================================================

1548936914 [2019-01-31 12:15:14], 30713,1 ['Cooler voltage too low, check fuses and wires']

1548936768 [2019-01-31 12:12:48], 30744,0

1548936768 [2019-01-31 12:12:48], 30739,3

Theoretically, below steps should be considered to check the problem
---Check the cables connecting the compressor and thermostat control board
---Check the cables between G7129A mainboard and thermostat control board
---Replacing the sample thermostat
---Replacing the G7129A mainboard
Unfortunately, the problem was still there after above trouble shooting.

A: Finally, we realized the compressor voltage shown in lab advisor is 12V instead of 24V which is normal. Then FSE checked the physical voltage supplied to thermostat, it turned out to be 24V. So, this is related to a signal processing issue electronically. At last, FSE changed the fusion board to resolve the issue.
Mar 15th 2019
Q: How to enable the G4260B to support Lan communication?
A:

---In order to enable the Lan communication of G4260B, you have to have
     a. update the FW to 32.06
     b. use latest ELSD driver 01.07(Waters provide the driver called ICS, the latest version is 3.0. Waters have to check whether ELSD driver 01.07 has been implemented in ICS 3.0 or not)
---For FW 32.06, you can only get .ELSD firmware, not Hex any more.
---To use .ELSD FW, you must use our labadvisor with latest ELSD Addon>>>>pls update your labadvisor and ELSD addon
---To use FW 32.06, you have to be sure that the bootloader of G4260B is 01.01.03 and above(if version not high enough, you got black screen after FW update)
---To check the bootloader version, you need to observe the first display of G4260B panel when it is powered on
---When updating FW, only series port is supported, no Lan
---When updating FW, you must specify the directory of folder in labadvisor , where you store your ELSD FW

---In order to enable the Lan communication of G4260B, you have to have
     a. update the FW to 32.06
     b. use latest ELSD driver 01.07(Waters provide the driver called ICS, the latest version is 3.0. Waters have to check whether ELSD driver 01.07 has been implemented in ICS 3.0 or not)
---For FW 32.06, you can only get .ELSD firmware, not Hex any more.
---To use .ELSD FW, you must use our labadvisor with latest ELSD Addon>>>>pls update your labadvisor and ELSD addon
---To use FW 32.06, you have to be sure that the bootloader of G4260B is 01.01.03 and above(if version not high enough, you got black screen after FW update)
---To check the bootloader version, you need to observe the first display of G4260B panel when it is powered on
---When updating FW, only series port is supported, no Lan
---When updating FW, you must specify the directory of folder in labadvisor , where you store your ELSD FW

Hi All,

Every time the modules show a solid red light at the front, means there is a hardware error, so this is immediately written into the firmware file of the same module. Our "softies" in Waldbronn spent loads of time writing error codes into the firmware and we can see these errors with the diagnostic buffers and system report.

So next time you see there is an error, instead of wasting your time describing it, taking snapshots from the GUI, looking for the software logbook, etc, it is better to obtain the diagnostic buffers and system report from Lab Advisor and attach them with you problem description.

Regards,

Ángel
July 26th 2017

Problem description: G7167A Initialization fail with error 25235,2 ['Transport motor index missing'] caused by incorrect installation of mainboard

Detail: When powering on the G7167A, the initial process begins successfully with below steps which means the FW needed for initialization was loaded without problem
          -- Blinking LED on fusion board
          --Changing status LED color
          --60s evacuation with high fan speed
          --Light is on in sampler housing
          However, when the sample handler is just beginning to move, we got the error message mentioned above. 
          Investigation was done by opening the cover of electronic box. We found the mainboard was installed incorrectly. In the front of the electronic box, the mainboard was installed under the metal rim which was designed to support the mainboard. The correct way should be the opposite, the board  should be put on the top of the metal rim, otherwise, you will create many short circuits on the mainboard. See below picture

Hello,

It is now possible to establish communication through LAN instead of RS232 in all G4260B ELSD. Default IP address for the ELSD is 192.168.254.27

Requirements:

-Minimum FW 32.06
-ELSD driver A.01.07
-Boot loader version 01.01.03 or ELSD with SN higher than GB1425XXXX. In case the boot loader version is inferior to 01.01.03, please check SN ELSD-017B to upgrade the boot loader version

Below is the ESLD drivers and Software versions compatibility. The main issue is that ELSD drivers A.01.07 are not compatible with ICF drivers or Chemstation rev B, so not LAN connectivity available with these software version,. For the rest of the software platforms, please check the minimum versions.













Hello,

In the Lab Advisor version B.02.11 there is a new test for the multisampler. It is the robot handler test and it is very useful to diagnose errors during initialization. With this test we can isolate problems on the robot and we always recommend it to run it before checking the belt tension or adjusting the belt tension screw. If the test is passed, we are 100% sure the robot is not faulty, so the problem must lie somewhere else. If the test fails we recommend to submit the report and lrs files to channel assist for evaluation.
Hi,

Many times the wavelength accuracy test fails after recalibrating the FLD with glycogen.

Here are some hints:

-Do not touch the monochromators assemblies. The don't need cleaning.
-Inspect the quartz window and replace it if the unit is too old. The window can generate a "milky spot" in the center due to ageing.
-Use latest FW and Lab Advisor versions
-Wait at least 20 minutes to run the wavelength accuracy test after recalibrating the detector. The flow cell may still contain traces of glycogen, so it must be thoroughly flushed with water.
-For G1321X versions use version Lab Advisor version B.02.12. At the time of publishing this article B.02.12 is not available to the field, but there is a B version that can be distributed to service engineers. Please don't provide to customers!!
New flow block
Part number for the new flow block G3488-81200

Original part was a softer metal prone to bending even during manufacturing. Known issue, No Service note. 




Part number for the RZZ mechanism for the H-Class SM-FTN is 700010179 and has to be ordered through EPO. Individual parts of this cannot be ordered, with the exception of the needle carriage.
Part number for the 2998 PDA Optical Bench is 700009328.
Part number for the beige PharMed needle wash tubing on the Acquity Classis SM is MV-W-430001002.
If you need to replace the on/off switch on the CM-A, you will need to order the entire door assembly as the switch is not sold separately. PN for the CM-A door is 700005443.
Part number for the Acquity CH-30A door hinge pin is WAT022186.
Part number for the SM-FTN needle wash tubing is 700005214

Part number for the SM-FTN purge tubing is 430002462
Part number for the fittings on the seal wash tubing inlet side are: peek nut, MV-W-700009441; ferrule with lock ring: MV-W-700009440. Both come as 3/pk.

Part number for the fittings on the seal wash tubing outlet side are: barbed fitting, 410001362.
Part number for the H-Class SM-FTN injection port T8 screw is MV-W-WAT081265.
When performing the Gradient test on a QUAT pump with AIV, there have been many reports of failures at either the 95%, 25% or both for the coefficient of determination.

Initially the models identified were G1311B/C and G7111A/B.  However, there have also been a few recent reports of G1311A also failing.  The failure is most likely to occur on the G1311A if either the MCGV or complete AIV were replaced.

The SU has conducted an investigation and have made improvements to the MCGV, AIV and updated firmware.  In addition, if a failure occurs, the SU has recommended the following steps.  We suggest you follow these steps BEFORE you replace any component.

1. Issue:
Some Quaternary 1260 pumps (G1311B/C) or 1260 Infinity II pumps (G7111A/B) are not
passing the OQPV Gradient Test. The investigation of returned parts showed that the
majority is working within specifications. Please take best precautions to avoid unnecessary
replacements.
2. Symptoms:
One or both of the following symptoms may be visible.
1. Composition accuracy (step gradient part) is failing.
2. Gradient Linearity is failing with excessive bumps at 95% tracer and 5% tracer.
3. Suggested repair strategy (future changes expected):
1. Make sure the pump has firmware D.07.13 or later loaded (important for correct STROKE
value in Auto Stroke mode)
2. Check the solvent intake filters. Replace them if required (not a problem for new pumps)
3. Thoroughly flush the system
1. Solvent: IPA in all channels
2. Solvent composition: 25 % each channel
3. Stroke volume 100 μL (for flushing only, set to AUTO for the OQ)
4. Flow rate: 3 mL/min
5. Duration: 10 min
4. Re-flush the system with water/tracer (make sure no air is getting into the flow path)
5. Select the appropriate Gradient Test
1. PIV: Use the combined Gradient Test (A/B/C/D)
2. AIV: Use the Dual Gradient Test (runs A/B and C/D separately)
6. Run the Gradient Test.
4. If the Gradient Test is still not passing:
1. Separate the 1260 Infinity II stack into two stacks. Put the solvent cabinet directly on top of
the pump module.
1. Start over with step 3.3
2. Permanently replace the AIV by a PIV (requires ok from customer!)
1. Start over with step 3.3.
3. Replace the MCGV. In rare occasions this may have to be more than once. Please take best
precautions to avoid unnecessary replacements.
1. Start over with step 3.3
5. Factory test situation (May 4, 2017)
1. The pump end test in WAD production does not include the OQPV gradient test.
2. For repair part MCGVs (PN G1311-67701) there is no chromatographic test nor OQPV test
done in WAD production.
3. In some recent cases it was found that properly assembled and adjusted valves did not
perform at customer site. One potential cause are shipment damages. The cardboard box
therefore being improved. Please handle MCGVs with care!
Problem:  When performing an autotune, when report is being generated the message pops up "mass offset is too high for this mass range".  Also when scanning below 10 the same error will come up.

Cause: Normal values for the mass offset for Single quad should be ~-10 to -33.  If user is getting higher positive number something in the mass filter electronics could be defective.

Solution:  Check the DIP values using the optimize quadrupole frequency in chemstation or Mass Hunter.  Normal values should be


If values for Dip at 10 AMU is much lower the 9, the software will show the error message.

Causes for the lower value have been attributed to a defective main board but could be cause by a problem with either quad driver (side board) or mass filter (gold quad , contactgs)

After replacing main board or other components, overwrite the autotune.u with a default copy.
PN for the I-Class SM-FL On/Off Switch is 441000555.
The PN for the Class Acquity BSM Personality board is MV-W-700009786/MV-X-700009786 and is compatible with both the passive check valve and i2V versions of the BSM. The OEM part is restricted and must be installed by a Waters FSE. There is an exchange part for this PCB.
PN for the H-Class Bio Flow Restrictor is 205000547. This can be used when additional backpressure is needed. It is 40um x 14" and is inert. Can be used with the CM-A, CH-A and CM-Aux.

July 2019

The ribbon cable of G7115A which connect optical unit to main board, it listed G1315-81600 in our service manual. It is actually too short and we do use 8121-2711 in our production

27 Nov 2017
Q: What are the basic steps to enhance the intensity of G1321X FLD?
A: The intensity is very important for Agilent FLD, without good intensity, the accuracy test will fail, the signal to noise test will fail as well.

1. The lamp. You may replace. See attached document on how to adjust
2. The quartz window. You may clean or replace

3. The EX condenser. You may replace

4. The alignment of the light beam on the EX slit. You need to check or adjust the light beam. See attached document on how to adjust
5. The mirror. You may replace

6. Finally the EX grating.
The expected intensity profile is like below

If all above replacements can not enhance the intensity, it is not worthwhile to do further replacement, pls change the optical unit.

Hi all,

Just for clarification a brief description of the different types of T sensors:

FY19 Oct 8th

Q: How does Chemstation LC driver detect the G1330B?

A: There three scenarios.

      Scenario one: the LC driver(Chemstation) already has a G1330B configured before. But one day, another G1330B with difference series number is installed and replace the original one

           >>>>>>No reconfiguration is needed, the driver does not check the series number of G1330B

    Scenario two: the LC driver(Chemstation) already has a G1330B configured before. But one day, the G1330B module is disconnected with ALS due to various reasons

           >>>>>>Reconfiguration is needed, otherwise the ALS will be offline when Chemstaion is opened, with prompt message "thermostat installation changed from true to false"

    Scenario three: the LC driver(Chemstation) has no G1330B configured before. But one day, a G1330B module is added into existing system.

           >>>>>>Reconfiguration is needed, otherwise the ALS will be offline when Chemstaion is opened, with prompt message "thermostat installation changed from false to true"

    

Error code from the front panel (External LED indicator) will provided a good idea on the issue:

0 R R 0 - 6 - X -Axis Collision Error
0 R R R - 7 - Theta-Axis Collision Error
R 0 0 0 - 8 - Z-Axis Collision Error

The way on how collision is generated will provide a very good idea of how to troubleshoot it, providing a video showing the collision will be very useful.

There are some typical parts to review evaluate for an specific Axis collision issue: FFC cable, Motor, Encoder wheel, belt, Sensor PCA and Main PCA.

1. Mechanical problem

- Verify first for any Axis  mechanical obstruction, check freely movement along all the way of mechanical parts
- Verify opto sensor conditions: dust accumulation, corrosive condition or misalignment, clean and/or re check opto alignment
- Double check on the belt condition: dry,  missing teeth or incorrect tension will cause this error
- Adjust belt tension, review service manual for details
- An Intermittent collision will suggest a mechanical issue
- Verify motor movement, if motor doesn't move, check voltage, if Voltage is  ok then change axis motor

Suggested parts for a mechanical related problem are:
- Axis Belt and/or related mechanical parts (pulleys)
- Axis motor
- Axis encoder PCB

2. Damage part / Communication problem

- Verify encoder alignment, if after encoder re alignment home position can't be find and/or opto can't be cleaned then an axis PCB is related.
- Double check on the FFC cable leads and connectors conditions, damage on lead and/or corrosive conditions will cause communication problems.
- If axis motor moves, but voltage is incorrect then an axis PCB, FFC cable or Main board problem is related.

Suggested parts for a mechanical related problem are:
- FFC cables
- Axis encoder PCB
- Main PCB

Follow instructions in  service manuals for details:

https://extranet.chem.agilent.com/sites/partners/support/Content%20Library/Agilent%20Internal%20and%20Entitled%20Partners/SPS_Service_Manual.pdf, also verify for any damage for stability on pulley mechanism.

https://extranet.chem.agilent.com/sites/partners/support/Content%20Library/Agilent%20Internal%20and%20Entitled%20Partners/G8410-90051.pdf

IMPORTANT: For SPS3 sold in 2014 or older, review SN LC A-030 for PCB updates on SPS3

https://extranet.chem.agilent.com/sites/partners/support/Service%20Notes/Service%20Note%20Viewer/LC/VLCA/SN_VLCA-030.htm

22 Dec 2017
Q: What are the part numbers for the tubing and connectors for Leak Waste of Infinity II are available for the field?
A: See below and attached document.

 

5500-1223

Tube Connector 90° (PK of 5)

5500-1221

Y Tube Connector (PK of 5)

5500-1222

T Tube Connector (PK of 5)

5500-1411

Tube Connector 180° (PK of 5)

5067-6137

E Tube Connector Kit (for pumps)

 

5043-1013            Tubing Clip IF-II

5067-5967            Tubing Clip Tube Connector (for MLS only)

5063-6527            Tubing for Leak Waste

FY19 Dec

Q: What is the part number for clamps to fix LC waste tubings at the side of LC module?

A:  I recommend you to order 5065-9976, this contains 10 1400-3676, and another 10 micro clamps .

15770630958afc653cdf393bd75098b65561bbd497854e4bd0.png



Best regards,



Steve Zhang

Global Service Channel Assist

----------------------------------------------------------------------------------------------------

Q: What is the part number of the fuses responsible for cooler/thermostat on Infinity II ALS mainboard?

A:  There are two fuses on the mainboard responsible for cooler/thermostat. The big one is with the shape of cylinder, it is a ceramic one, the part number is 2110-1519. The other one is a small square shape fuse, the part number is 2110-1519. 

Jan 2020: The serial number is stored in the dynamicParms.txt file, located in the /home/sc4/pipeLineQTOF directory of the smartcard.

 

  • Below is the content of an example dynamicParms file(*). You will need to create this file to the correct serial number and save it in C:\tmp You will need to create this directory.

 

  • Stop the firmware on the instrument – telnet to the backdoor and enter: stopfw

 

  • Next, you FTP the dynamicParms.txt file to the smartcard.
    1. Start the command line editor from run: CMD
    2. In CMD, change to the  C:\tmp folder: CD C:\tmp
    3. FTP into the SC: ftp 192.168.254.12 (you will need to log in:  UN: sc4 and PW: sc4demo)
    4. Change to the pipeLineQTOF directory: cd pipeLineQTOF
    5. Change to binary mode: bin
    6. Transfer the file: put dynamicParms.txt
    7. Quit ftp: bye

 

  • Power cycle the instrument and once booted, check the backdoor for serial number update: showrevs

Create this file:

dynamicParms.txt

1.000,0,US1202H001,QTOF

Feb 2020

597x CI Methane CH4 flow adjustment fails for 27/28 ratio or is very high

 

If 27 is higher than 28 most of the cases its an issue related to the CH4 pressure within the CI ion source.

In this case the ratio for 27/28 with a Ch4 flow of 20% was at about 0.5 and it required a CH4 flow of ~90% to correct this.

 

Troubleshooting:

  • CH4 pressure at the unit was around 15 psi
  • Gas quality was ok, 5.5 and filters been installed properly
  • CI transferline tip was installed
  • Last time the system was properly tuning in CI mode was 10 years ago

 

Solution:

  • The wrong ceramic tip was installed at the new style transferline, so the CI source was not tight enough.

Problem:

Issue and solution:

 